Q

Examination vehicle, ball tube and image intensifier deviate up and down

A
1. Check the microswitch on the side of the frame of the ball tube and image intensifier 2. Open the electrical control cabinet to check whether the wire ends are loose
Q

The CK312A host reports error 5

A
1. Check whether the 125v/10A insurance in the insurance works normally 2. Check whether the filament plate works normally
Q

UC arm crossarm cannot be up or down or rotated

A
1. Enter the program and check whether it can work in the program 2. If it can work in the program, it is necessary to reset the motion parameters 3, if in the program can not work, remove the case to check the insurance on the machine
Q

The DR Enters the DROC software and reports License node restriction does not match client,s node for product DROC work in DEMO mode.

A
Go to drive C and open the DROC folder to look for the watch Dog Detelt file. Open the file and click Evalvetioninfo to check the time limit and contact our technical staff
Q

The DR Message "panel initialization failed, please check if configuration is correct" is displayed.

A
Possible causes: 1. Check whether the plate is energized 2. Check whether the network connection between the tablet and the computer host is properly connected